Description
SPIROBANK II SMART + WINSPIROPRO SOFTWARE + OXYMETER
OXYMETER: more than 300 hours of oximetric test records (SpO2 % and pulse frequency tests).
Measured parameters: min, max and average SpO2 %, min, max and average pulse frequency, SpO2 total events, test duration, bradycardia events (<40 BPM), tachycardia events (>120 BPM), T90% (time passed with SpO2 =89%), T89% (time passed with SpO2 = 88%).
Spirometer with 10.000 tests memory capacity, monochromatic LCD display, 6 keys membrane keypad. Battery power supply (3,7 V-1100 mA lithium) with an autonomy of approx. 50 hours of continuous use, rechargeable with USB port.
- cable (USB 2.0) or wireless (Bluetooth 4.0) connection
- internal temperature sensor for automatic BTPS conversion
- high resolution graphic display
- automatic record of a whole spirometry session including best test
- USB cable, carrying case, CD manual (GB, FR, IT, ES, DE, PT) re-usable turbine and software included. Selectable internal software (GB, FR, IT, ES, DE, PT)
WinspiroPRO software functions: (PC system requirements: XP, Seven, Vista, 8, 10)
- Languages: GB, FR, IT, DE, ES, PT, PL, SE, NL, LV, CZ, TR, RU, CN, JP
- on-line PC connection with icon interface
- real time Flow/Volume and Volume/time curves
- bronchial challenge with FEV1 dose-response
- paediatric incentive animations
- integration with EPR systems
- Lung Age Estimation
- data and graphs export also via e-mail
PC system requirements: XP, Seven, Vista, 8, 10
160 x 80 pixel display
Weight 140 g
Size 55 x 160 x 25 mm
PC connection via USB or Bluetooth®
Compatible with iPad or mini iPad tablets
Technical Specifications
TECHNICAL SPECIFICATIONS - SPIROMETRY
• Measured parameters: FVC, FEV1, FEV1/FVC%, DTPEF, FEV 0.5, FEV0.5/FVC%, FEV0.75, FEV0.75/ FVC%, FEV2, FEV2/FVC%, FEV3, FEV3/FVC%, FEV6, FEV1/FEV6%, PEF, FEF25%, FEF50%, FEF75%, FEF25-75%, FEF75-85%, FET, Vext, Lung Age, EVOL, FIVC, FIV1, PIF, FIV1/FIVC%, FIF25%, FIF50%, FIF75%, R50, PIF, IRV, VC, IVC, IC, ERV, FEV1/VC%, VT, VE, Rf, ti, te, ti/t-tot, VT/ti, MVV
• Flow sensor bi-directional digital turbine
• Flow range ±16L/s
• Volume accuracy ±3% or 50 mL
• Flow accuracy ±5% or 200 mL/s
• Dynamic resistance <0.5 cm H2O/L/s
• Available tests: FVC, VC, IVC, MVV, PRE - POST
• Connectivity: USB 2.0, Bluetooth® 4.0
• Power supply: Lithium-Ion 3.7V, 1100 mAh
TECHNICAL SPECIFICATIONS - OXIMETRY
• Measurement method: red and infrared absorption
• SpO2 range: 0-99% ±2 % between 70-99% SpO2
• Pulse rate range: 18-300 BPM ±2 BPM or 2%
• Signal quality indicator: 0-8 segments on display
• Memory capacity: up to 300 hours oximetry